Skip to main content

Table 5 Patient-averaged values of total monitor units (MUs) and MU reduction with respect to the optimal collimator angles, and p values for one way analysis of variance (ANOVA) with Bonferroni multiple comparison of the patient-averaged values of MU

From: Optimal collimator rotation based on the outline of multiple brain targets in VMAT

 

Std-VMAT

Colli-VMATPre (40°)

Colli-VMATPre (60°)

Colli-VMAT (40°)

Colli-VMAT (60°)

Average

472 ± 235

400 ± 154

419 ± 168

390 ± 148

417 ± 167

MU reduction (%)

0.0

−15.4

−11.3

−17.4

−11.8

Bonferroni adjusted p-values

 Colli-VMATPre (40°)

0.004

 Colli-VMATPre (60°)

0.011

0.029

 Colli-VMAT (40°)

0.010

0.012

0.046

 

 Colli-VMAT (60°)

0.023

0.057

0.356

0.047