Skip to main content

Table 1 The scan time and reconstruction slice number at 6 groups of scan options on TomoTherapy MVCT image-guided radiotherapy system

From: Effects of megavoltage computed tomographic scan methodology on setup verification and adaptive dose calculation in helical TomoTherapy

Scan Options

Chest Phantom

Pelvic Phantom

Acquisition Pitch

Reconstruction Interval

Scan Length (mm)

Scan Time (s)

Reconstruction Slices

Scan Length (mm)

Scan Time (s)

Reconstruction Slices

Fine

(4 mm/ra)

1 mm

135

356

135

106

283

106

2 mm

137

361

69

106

281

53

Normal

(8 mm/r)

2 mm

136

191

69

106

151

53

4 mm

134

186

34

107

151

27

Coarse

(12 mm/r)

3 mm

139

136

47

108

108

36

6 mm

137

131

23

114

111

19

  1. amm/r: the movement distance of treatment couch during one rotation of the beam gantry