Skip to main content

Table 2 Target volume dose distribution

From: Comparison of photon volumetric modulated arc therapy, intensity-modulated proton therapy, and intensity-modulated carbon ion therapy for delivery of hypo-fractionated thoracic radiotherapy

Parameters

VMAT1

IMPT2

IMCIT3

 

p value

 

Mean (SD)

p 1,2

p 2,3

p 1,3

V95

100% (0)

99% (1%)

100% (0)

1.00

0.02

0.46

D95 a

59.57 (0.40)

59.63 (0.67)

59.85 (0.47)

0.17

0.04

0.01

D99 a

58.00 (0.75)

57.99 (1.25)

58.38 (0.81)

0.46

0.02

0.21

Dmax a

64.87 (0.92)

65.98 (1.14)

65.25 (0.77)

0.04

0.01

0.41

CI

1.25 (0.10)

1.37 (0.11)

1.30 (0.09)

0.01

0.00

0.11

HI

1.08 (0.02)

1.10 (0.02)

1.09 (0.01)

0.03

0.02

0.20

nCI

1.47 (0.80)

1.53 (0.78)

1.47 (0.77)

0.01

0.00

0.79

  1. p 1,2 is VMAT vs. IMPT; p 2,3 is IMPT vs. IMCIT; p 1,3 is VMAT vs. IMCIT. aDose in Gy