Table 2 Comparison of delivery

From: Simultaneous integrated boost therapy of carcinoma of the hypopharynx/larynx with and without flattening filter - a treatment planning and dosimetry study

Parameter

IMRT FF

IMRT FFF

VMAT FF

VMAT FFF

Passing Rate of Gamma Evaluation (%)

98.5 ± 1.3

97.6 ± 2.0

98.6 ± 0.9

98.9 ± 0.9

Delivery Time (min:s)

9:27 ± 0:34

9:49 ± 0:28

2:52 ± 0:07

2:53 ± 0:05

Measured Peripheral Dose 30 frac (mGy)

209 ± 36

170 ± 34

155 ± 42

127 ± 38

Monitor Units

859 ± 70a

1181 ± 120

571 ± 63a

694 ± 59

Segments (IMRT)

71 ± 4a

79 ± 3

-

-

  1. Mean values and standard deviation of delivery time, monitor units and segments for FF and FFF mode averaged over all patients. Bold values indicate statistically significant superior values in the comparison of FF vs FFF. Bold-italic values indicate best values in the comparison of all planning techniques and irradiation modes. a indicates statistical significance in the comparison of FF vs FFF without judgement, since the number of monitor units and the number of segments are no measures of quality