Skip to main content

Table 2 target volume objectives for dose coverage

From: Primary tumor volume delineation in head and neck cancer: missing the tip of the iceberg?

Objectiveb

Per Protocol

Variation Acceptable

Deviation Unacceptable

D95% of PTV70

70 Gy (100%)

≤+/-2%

>+/-2%

D5% of PTV70

≤74.9 Gy (107%)

≤77 Gy (110%)

>77 Gy (110%)

D98% of PTV70

≥66.5 Gy (95%)

≥63 Gy (90%)a

<63 Gy (90%)a

Maximum dose (hot spot ≤0.03 cm3) outside the PTV70

≤74.9 Gy (107%)

≤77 Gy (110%)

>77 Gy (110%)

  1. aIt is recognized that portions of the PTV close to the skin may receive significantly less than 90% of the prescribed dose. This is acceptable as long as cold spots within PTV do not exist at a depth deeper than 5 mm beneath the skin
  2. bCTV_S and CTV_L used instead of PTV70 for the purpose of this study