Table 1 Camera setting Templates

From: Evaluation of daily patient positioning for radiotherapy with a commercial 3D surface-imaging system (Catalyst™)

Thorax

 Tolerance Settings

  lateral

5.0 mm

  longitudinal

5.0 mm

  vertical

5.0 mm

  rotation

5.0°

  roll

5.0°

  pitch

5.0°

 Surface Settings

  Tolerance

10 mm

 Image Surface Averaging

  Time

3 s

Abdomen

 Tolerance Settings

  lateral

5.0 mm

  longitudinal

5.0 mm

  vertical

5.0 mm

  rotation

5.0°

  roll

5.0°

  pitch

5.0°

 Surface Settings

  Tolerance

15 mm

 Image Surface Averaging

  Time

4 s

Pelvis

 Tolerance Settings

  lateral

8.0 mm

  longitudinal

8.0 mm

  vertical

8.0 mm

  rotation

5.0°

  roll

5.0°

  pitch

5.0°

 Surface Settings

  Tolerance

15 mm

 Image Surface Averaging

  Time

3 s

  1. Predefined templates for camera settings with tolerance values regarding the depiction of surface deviations and surface averaging time