Table 2 Comparison of plan delivery

From: Simultaneous integrated boost (SIB) radiation therapy of right sided breast cancer with and without flattening filter - A treatment planning study

Plan Time [min:sec] MU Control points
IMRT FF 09:29 ± 1:25 630 ± 114a 188 ± 31a
IMRT FFF 10:13 ± 1:30 792 ± 128 228 ± 39
VMAT FF 2:38 ± 0:17 713 ± 77a 162 ± 10a
VMAT FFF 2:27 ± 0:14 852 ± 105 181 ± 7
tVMAT FF 1:58 ± 0:14 651 ± 124a 53 ± 2
tVMAT FFF 1:20 ± 0:07 763 ± 96 52 ± 2
  1. Mean values and standard deviation of delivery time, monitor units and control points for FF and FFF mode averaged over all patients separated by the treatment technique. Bold values indicate statistically significant superior values in the comparison of FF vs FFF. Bold-italic letters indicate best values in the comparison of all planning techniques and irradiation modes. aindicates statistical significance in the comparison of FF vs FFF without judgement, since the number of monitor units and the number of control points are no measures of quality