Table 5 Results of error detection and classification

From: Investigation of a real-time EPID-based patient dose monitoring safety system using site-specific control limits

Treatment site Patient no. Number of fractions operated with Watchdog Overall % under control limit % Under condition of error detection
Clinical Non-clinical
Patient related delivery errors Transfer and machine errors Watchdog system errors Acquisition errors Watchdog user errors
Prostate 1 31 4.4 % 1.9 % 0.0 % 1.5 % 0.3 % 0.8 %
2 30 6.9 % 3.7 % 0.0 % 1.8 % 0.1 % 1.2 %
3 30 4.9 % 1.7 % 0.0 % 2.7 % 0.1 % 0.5 %
4 35 6.1 % 2.5 % 0.0 % 2.1 % 0.3 % 1.3 %
5 31 5.6 % 1.6 % 0.0 % 3.4 % 0.1 % 0.4 %
Head and neck 1 29 2.9 % 1.8 % 0.0 % 0.6 % 0.3 % 0.2 %
2 32 23.7 % 11.2 % 0.0 % 8.5 % 2.5 % 1.5 %
3 33 0.0 % 0.0 % 0.0 % 0.0 % 0.0 % 0.0 %
4 27 7.1 % 0.9 % 0.0 % 4.7 % 0.1 % 1.4 %
5 28 9.7 % 6.8 % 0.0 % 1.4 % 0.6 % 0.8 %
Rectum 1 20 0.0 % 0.0 % 0.0 % 0.0 % 0.0 % 0.0 %
2 14 13.3 % 7.9 % 0.0 % 2.0 % 1.8 % 1.6 %
3 25 4.9 % 1.3 % 0.0 % 1.8 % 0.2 % 1.6 %
4 24 7.0 % 1.3 % 0.0 % 3.8 % 0.2 % 1.6 %
5 29 8.4 % 4.3 % 0.0 % 2.9 % 0.1 % 1.2 %