Table 3 Descriptive statistics of the paired measurements

From: Implementation of an intraoperative electron radiotherapy in vivo dosimetry program

% Absorbed dose

MOSFET

Film

Minimum

78.0

72.0

1st quartile

90.2

92.4

Median

92.5

97.3

2nd quartile

97.6

103.0

Maximum

111.6

123.4

Mean

93.9

97.9

Standard deviation

6.8

9.7

  1. Thirty pairs of values for which the prescribed dose was 90 % and the maximum absorbed dose delivered was 100 %