Fig. 2From: A model for preemptive maintenance of medical linear accelerators—predictive maintenanceLinear accelerator predictive maintenance dashboard (PMD) illustrating a beam steering parameter (uniformity) change that was detected by the PdM process and confirmed via scanning water phantom. Continuous monitoring following service adjustments indicates beam steering was restored to baseline valueBack to article page