Table 1 PTV dose differences in lung SBRT between the CT datasets

From: Dosimetric impact of different CT datasets for stereotactic treatment planning using 3D conformal radiotherapy or volumetric modulated arc therapy

  

3D-CRT

VMAT

 

CT datasets

Δ [%]

P value

Δ [%]

P value

ΔDmean

PCT vs AIP

−0.8 ± 0.5

0.01*

−0.5 ± 0.5

0.01*

 

MIP vs AIP

1.5 ± 1.4

0.01*

1.6 ± 1.5

0.01*

 

MidV vs AIP

0.0 ± 0.3

0.96

−0.1 ± 0.3

0.72

 

MIP vs PCT

2.4 ± 0.7

0.01*

2.2 ± 1.5

0.01*

 

MidV vs PCT

0.8 ± 0.7

0.02*

0.5 ± 0.6

0.06

 

MIP vs MidV

1.6 ± 1.3

0.01*

1.7 ± 1.5

0.01*

ΔD2

PCT vs AIP

−0.5 ± 0.7

0.04*

−0.2 ± 0.8

0.29

 

MIP vs AIP

1.0 ± 1.5

0.07

1.6 ± 2.0

0.01*

 

MidV vs AIP

−0.1 ± 0.4

0.39

0.0 ± 0.5

0.96

 

MIP vs PCT

1.5 ± 2.0

0.01*

1.9 ± 1.9

0.01*

 

MidV vs PCT

0.4 ± 0.9

0.13

0.2 ± 1.1

0.33

 

MIP vs MidV

1.1 ± 1.6

0.06

1.7 ± 2.0

0.01*

ΔD98

PCT vs AIP

−1.4 ± 0.4

0.01*

−1.1 ± 0.7

0.01*

 

MIP vs AIP

1.0 ± 1.6

0.07

1.0 ± 1.4

0.02*

 

MidV vs AIP

0.2 ± 0.4

0.09

0.0 ± 0.4

0.59

 

MIP vs PCT

2.4 ± 1.9

0.01*

2.2 ± 1.7

0.01*

 

MidV vs PCT

1.7 ± 0.6

0.01*

1.1 ± 0.7

0.01*

 

MIP vs MidV

0.7 ± 1.5

0.20

1.0 ± 1.4

0.07

ΔD95

PCT vs AIP

−1.3 ± 0.3

0.01*

−0.9 ± 0.4

0.01*

 

MIP vs AIP

1.2 ± 1.6

0.02*

1.0 ± 1.4

0.03*

 

MidV vs AIP

0.2 ± 0.4

0.09

0.0 ± 0.4

0.68

 

MIP vs PCT

2.5 ± 1.8

0.01*

1.9 ± 1.6

0.01*

 

MidV vs PCT

1.5 ± 0.5

0.01*

0.9 ± 0.4

0.01*

 

MIP vs MidV

1.0 ± 1.4

0.04*

1.0 ± 1.5

0.06

  1. Relative differences over all patients (mean ± standard deviation) of dosimetric values for PTV in lung SBRT between the different CT datasets for 3D-CRT and VMAT plans. PCT planning CT, AIP average intensity projection CT, MIP maximum intensity projection CT, MidV mid-ventilation CT. *assigns statistical significant values (p < 0.05)