Fig. 3 | Radiation Oncology

Fig. 3

From: Current state of the art of regional hyperthermia treatment planning: a review

Fig. 3

- Analysis of heating effectiveness by HTP. Treatment planning example demonstrating the influence of operating frequency and the number of antennas and rings on the heating effectiveness of a cervical cancer patient. The picture on the left shows the patient anatomy and the other pictures show the simulated temperature distributions for heating with a double ring system at 70 MHz or 130 MHz and a 130 MHz triple ring system. Both higher frequency and increasing number of antennas improve the steering properties and heating effectiveness. The simulated T90 was 40.3, 40.9 and 41.9 °C for the three cases, respectively. This figure was reproduced from Kok HP, De Greef M, Borsboom PP, Bel A, Crezee J: Improved power steering with double and triple ring waveguide systems: the impact of the operating frequency. Int. J. Hyperthermia 2011, 27:224–239, copyright © 2011, Informa Healthcare. Reproduced with permission of Informa Healthcare