Table 4 Comparison of D90 and D95 between APBI devices

From: Comparative dosimetric findings using accelerated partial breast irradiation across five catheter subtypes

 

(I) Applicator

(J) Applicator

Mean Difference (I–J)

Std. Error

p

D90

Single-lumen

Multi-lumen

3.1699

1.04264

.002

Savi

−4.1677

.89960

< .001

Savi-mini

−3.1167

.89741

.001

Multi-lumen

Single-lumen

−3.1699

1.04264

.002

Savi

−7.3376

.69028

< .001

Savi-mini

−6.2866

.68743

< .001

Savi

Single-lumen

4.1677

.89960

< .001

Multi-lumen

7.3376

.69028

< .001

Savi-mini

1.0510

.44129

.018

Savi-mini

Single-lumen

3.1167

.89741

.001

Multi-lumen

6.2866

.68743

< .001

Savi

−1.0510

.44129

.018

D95

Single-lumen

Multi-lumen

5.6730

1.29051

< .001

Savi

−.3506

1.11346

.753

Savi-mini

1.6171

1.11076

.146

Multi-lumen

Single-lumen

−5.6730

1.29051

< .001

Savi

−6.0236

.85439

< .001

Savi-mini

−4.0559

.85086

< .001

Savi

Single-lumen

.3506

1.11346

.753

Multi-lumen

6.0236

.85439

< .001

Savi-mini

1.9677

.54620

< .001

Savi-mini

Single-lumen

−1.6171

1.11076

.146

Multi-lumen

4.0559

.85086

< .001

Savi

−1.9677

.54620

< .001