Table 3 Comparison of skin and rib doses between APBI devices

From: Comparative dosimetric findings using accelerated partial breast irradiation across five catheter subtypes

Dependent Variable

(I) Applicator

(J) Applicator

Mean Difference (I–J)

Std. Error

p

DmaxSkin

Single-lumen

Multi-lumen

−5.5242

6.20960

.374

Savi

27.5618

5.35769

< .001

Savi-mini

12.3611

5.34468

.021

Multi-lumen

Single-lumen

5.5242

6.20960

.374

Savi

33.0860

4.11108

< .001

Savi-mini

17.8853

4.09411

< .001

Savi

Single-lumen

−27.5618

5.35769

< .001

Multi-lumen

−33.0860

4.11108

< .001

Savi-mini

−15.2007

2.62820

< .001

Savi-mini

Single-lumen

−12.3611

5.34468

.021

Multi-lumen

−17.8853

4.09411

< .001

Savi

15.2007

2.62820

< .001

DmaxRib

Single-lumen

Multi-lumen

8.9750

7.52486

.233

Savi

35.9376

6.49250

< .001

Savi-mini

36.3416

6.47673

< .001

Multi-lumen

Single-lumen

−8.9750

7.52486

.233

Savi

26.9626

4.98185

< .001

Savi-mini

27.3666

4.96129

< .001

Savi

Single-lumen

−35.9376

6.49250

< .001

Multi-lumen

−26.9626

4.98185

< .001

Savi-mini

.4040

3.18487

.899

Savi-mini

Single-lumen

−36.3416

6.47673

< .001

Multi-lumen

−27.3666

4.96129

< .001

Savi

−.4040

3.18487

.899