Skip to main content

Table 1 Comparing prostate intrafraction motion magnitude. Prostate intrafraction motion results from multiple institutional trials using an endorectal balloon (ERB) for immobilization compared to the current study using an EIS. Patients treated with the EIS showed significantly reduced 3D intrafraction motion for treatment times longer than 6 min. EM = Electromagnetic

From: Clinical evaluation of an endorectal immobilization system for use in prostate hypofractionated Stereotactic Ablative Body Radiotherapy (SABR)

Study

Sample size (n)

Technique

Immobilization

Motion Evaluation

3D intrafraction motion

Treatment time (min)

Current study

20

VMAT

EIS

Pre-post CBCT

0 % > 3 mm

>6

Chiang et al. 2014 [36]

30

VMAT

ERB

Pre-post CBCT

3.7 % > 4.5 mm

>6

Smeenk et al. 2012 [15]

30

IMRT

ERB

Real-time EM tracking

7.0 % > 3 mm

<6

Wang et al. 2012 [34]

30

IMRT or VMAT

ERB

Real-time EM tracking

5.0 % > 3 mm

 ≤6

Both et al. 2011 [14]

24

IMRT or VMAT

ERB

Real-time EM tracking

5.2 % > 3 mm

<6