Table 2 Farmer ionization chamber readings without (M) and with TachoSil (M T )

From: Attenuation measurements show that the presence of a TachoSil surgical patch will not compromise target irradiation in intra-operative electron radiation therapy or high-dose-rate brachytherapy

M (nC) M T (nC) Dif T
3.486 ± 0.001 3.469 ± 0.001 −0.5%