Table 2 Statistical comparison of the tested linacs for the average and largest range of deviations, and the maximum standard deviations of the results (worst reproducibility) over each of the tested components

From: An EPID-based method for comprehensive verification of gantry, EPID and the MLC carriage positional accuracy in Varian linacs during arc treatments

  

Average range

Largest range

Worst reproducibility

EPID Sag (mm)

Cross-plane

0.17

0.32 (L#11)

0.04 (L#11)

In-plane

1.00

2.68 (L#7)

0.08 (L#10)

Gantry Sag (mm)

Cross-plane

0.42

0.63 (L#11)

0.10 (L#10)

In-plane

0.72

0.82 (L#12)

0.15 (L#8)

SDD (mm)

 

3.57

8.65 (L#11)

1.15 (L#2)

EPID and Collimator Skew (degrees)

 

0.15

0.17 (L#9)

0.09 (L#3)

EPID Tilt (degrees)

Cross-plane

0.02

0.04 (L#10)

0.01 (L#7)

In-plane

0.06

0.13 (L#11)

0.05 (L#11)

MLC Sag (mm)

Left

0.62

0.82 (L#8)

0.17 (L#3)

Right

0.68

0.99 (L#8)

0.15 (L#3)

Gun

0.34

0.57 (L#11)

0.16 (L#6)

Target

0.32

0.61 (L#10)

0.10 (L#3)

  1. Values are in mm except for the EPID tilt and skew which are given in degrees. The largest measured values are listed and their corresponding linac numbers are given in parentheses.