Skip to main content

Table 2 MOSFET dose measurement 4 MeV TSEI for dosimetric commissioning and treatment MUs

From: Dosimetric comparison of 4 MeV and 6 MeV electron beams for total skin irradiation

Detector location 1stday dose (cGy) 2ndday dose (cGy) Total dose (cGy) Normalized dose (cGy) Dose difference (%)
ANT of Chest 110 201 311 200.8 0.4
Right of Chest 140 168 308 198.9 -0.6
POST of Chest 197 103 300 193.7 -3.1
Light of Chest 135 160 295 190.5 -4.8
ANT of Umbilicus 115 221 336 216.9 8.5
Right of Umbilicus 154 161 315 203.4 1.7
POST of Umbilicus 203 104 307 198.2 -0.9
Light of Umbilicus 151 155 306 197.6 -1.2
Average    309.8 200.0 0.0
  1. First day beams: AP, LPO, and RPO Second day beams: PA, LAO, and RAO.