From: Dosimetric comparison of 4 MeV and 6 MeV electron beams for total skin irradiation
Detector location | 1stday dose (cGy) | 2ndday dose (cGy) | Total dose (cGy) | Normalized dose (cGy) | Dose difference (%) |
---|---|---|---|---|---|
ANT of Chest | 110 | 201 | 311 | 200.8 | 0.4 |
Right of Chest | 140 | 168 | 308 | 198.9 | -0.6 |
POST of Chest | 197 | 103 | 300 | 193.7 | -3.1 |
Light of Chest | 135 | 160 | 295 | 190.5 | -4.8 |
ANT of Umbilicus | 115 | 221 | 336 | 216.9 | 8.5 |
Right of Umbilicus | 154 | 161 | 315 | 203.4 | 1.7 |
POST of Umbilicus | 203 | 104 | 307 | 198.2 | -0.9 |
Light of Umbilicus | 151 | 155 | 306 | 197.6 | -1.2 |
Average | 309.8 | 200.0 | 0.0 |