Skip to main content

Table 4 Area under curve (AUC) values for each gamma criterion when introducing 0.5 mm MLC misalignment in class in and class out and 2 mm MLC misalignment in class shift

From: The sensitivity of gamma-index method to the positioning errors of high-definition MLC in patient-specific VMAT QA for SBRT

 

MapCHECK2

EBT2 film

Gamma criterion

Class In (0.5 mm)

Class Out (0.5 mm)

Class Shift (2 mm)

Class In (0.5 mm)

Class Out (0.5 mm)

Class Shift (2 mm)

1%/1 mm

0.909

0.899

0.935

0.853

0.933

0.905

1.5%/1.5 mm

0.920

0.873

0.974

0.788

0.858

0.89

1%/2 mm

0.843

0.890

0.954

0.758

0.813

0.805

2%/1 mm

0.930

0.848

0.968

0.865

0.925

0.933

2%/2 mm

0.883

0.719

0.958

0.783

0.824

0.873