Table 4 Mean errors of p CT and estimated mean deviations of WET for different tissue types

From: MRI-based treatment plan simulation and adaptation for ion radiotherapy using a classification-based approach

Tissue

Mean error of

Δ WET1cm

Δ WET5cm

Δ WET10cm

 

p CT [HU]

[mm]

[mm]

[mm]

air

43.3

0.4 ± 0.3

1.8 ± 0.6

3.6 ± 0.9

“partial volume”

28.6

0.3 ± 0.7

1.4 ± 1.5

2.9 ± 2.1

soft tissue

-1.4

-0.02 ± 0.3

-0.1 ± 0.6

-0.2 ± 0.9

soft bone

-63.2

-0.4 ± 0.3

-1.8 ± 0.6

-3.7 ± 0.9

bone

-223.1

-1.2 ± 0.4

-6.0 ± 0.8

-11.9 ± 1.1