Table 1 3D global gamma analysis of the reference plan with static beams

From: Comparison of dose calculations between pencil-beam and Monte Carlo algorithms of the iPlan RT in arc therapy using a homogenous phantom with 3DVH software

Gamma 3%/3 mm

Gamma passing rate (%)

p value

PBC

MCC

Target ROI

100 ± 0.00

100 ± 0.00

1.000

Periphery ROI

98.00 ± 0.10

99.70 ± 0.00

0.037

Entire area

98.10 ± 0.10

99.70 ± 0.00

0.034

Gamma 2%/2 mm

PBC

MCC

p value

Target ROI

100 ± 0.00

100 ± 0.00

1.000

Periphery ROI

94.20 ± 0.30

97.77 ± 0.12

0.046

Entire area

94.47 ± 0.31

97.90 ± 0.10

0.050

  1. Abbreviations: PBC Pencil-beam calculation, MCC Monte Carlo calculation.
  2. Two gamma criteria (3%/3 mm and 2%/2 mm) with a threshold of 5% of maximal dose were used. The 3D gamma passing rate of the entire area and of each region of interest (target and periphery ROI) obtained through the pencil-beam calculation (PBC) are compared with those obtained through the Monte Carlo calculation (MCC) using the Mann–Whitney U-test. The dose was measured three times, and is shown as mean ± the standard deviation.