Table 1 Results of the TLD measurements at the four positions in 3 mm depth (3.1 – 3.4) and 1 mm depth (O.1 – O.4) in percentage of 2 Gy (prescribed dose per fraction) for the 4 MV and the 6 MV treatment plan

From: Measurement of skin and target dose in post-mastectomy radiotherapy using 4 and 6 MV photon beams

Position 4 MV 6 MV (40/60)
O.1 69.4 84.1
O.2 70.5 76.7
O.3 73.1 84.7
O.4 78.7 81.0
3.1 95.1 95.7
3.2 94.6 95.2
3.3 93.1 94.8
3.4 94.9 93.0
  1. The 6 MV treatment without/with bolus was weighted 40/60.