From: In vitro irradiation system for radiobiological experiments

Figure 4

Dosimetric evaluation of calculated and measured dose distributions. (a) Comparison between the calculated and measured dose inside the recess holding the microplate at SSD 99 cm, evaluated using radiochromic films (Gafchromic EBT Films, International Specialty Products). (b) Planar distribution of gamma index representing the 98.4 pixels passing the test with dose tolerance of 5%. (c, d) Comparison between measured (red line) and calculated (blue line) dose profiles along the short and the long side, respectively, of the cut Gafchromic film. All images were created by Open Source DoseLab 4.00 software (http://doselab.sf.net).