Table 1 Patient specific IMRT QA absolute dose measurement analysis

From: A systematic approach to statistical analysis in dosimetry and patient-specific IMRT plan verification measurements

Meas. Num (n)

Measurement result (Gy)

Mean value (1 to n)

Result standard deviation (1 to n)

Estimated dose delivery standard deviation

Probability for the mean to be within 2% accuracy (1 to n)

Probability for the mean to be within 3% accuracy (1 to n)

a. Chamber 1: corresponding plan dose was 1.702 Gy

1

1.706

1.706

-

 

-

-

2

1.695

1.700

0.008

0.010

89.63%

93.05%

3

1.695

1.699

0.006

0.007

98.85%

99.50%

4

1.689

1.696

0.007

0.008

99.76%

99.93%

5

1.689

1.695

0.007

0.007

99.96%

99.98%

b. Chamber 2: corresponding plan dose was 0.546 Gy

1

0.513

0.513

-

 

-

-

2

0.507

0.510

0.005

0.006

80.50%

86.76%

3

0.504

0.508

0.005

0.006

93.11%

96.77%

4

0.504

0.507

0.005

0.005

97.94%

99.33%

5

0.502

0.506

0.005

0.005

99.25%

99.83%

  1. The measurements were conducted without the QA phantom re-setup between each measurement and the statistical analysis was for the accuracy of measurement itself. The accuracy is defined as percent difference between the expected value and measurement mean value.