Table 6 Gamma agreement index

From: Critical appraisal of the accuracy of Acuros-XB and Anisotropic Analytical Algorithm compared to measurement and calculations with the compass system in the delivery of RapidArc clinical plans

Group

AAA_Acuros

CC_Acuros

CC_AAA

p

GAI [%] with criteria 3 %/3 mm

 Brain

99.1 ± 2.1

99.2 ± 2.1

99.7 ± 0.6

b

 H&N

99.0 ± 1.4

99.4 ± 0.9

99.4 ± 1.0

a,b

 Thorax

99.6 ± 0.8

99.9 ± 0.2

99.7 ± 0.33

-

 Pelvis

99.4 ± 1.4

99.9 ± 0.3

98.9 ± 3.1

a,c

 SBRT

99.6 ± 1.4

100.0 ± 0.0

99.5 ± 2.1

-

GAI [%] with criteria 2 %/2 mm

 Brain

97.0 ± 4.5

98.0 ± 3.6

98.7 ± 3.4

a,b

 H&N

91.1 ± 7.3

96.9 ± 3.2

96.2 ± 5.5

a,b

 Thorax

96.4 ± 5.7

99.2 ± 1.3

97.9 ± 2.4

a,c

 Pelvis

92.2 ± 11.3

98.8 ± 2.3

91.9 ± 14.4

a,c

 SBRT

97.8 ± 4.8

99.9 ± 0.2

98.5 ± 4.9

a

  1. a AAA_Acuros vs CC_ Acuros, b AAA_ Acuros vs CC_AAA, c CC_ Acuros vs CC_AAA.