Table 2 Calculated [[24]] and measured carbon ion ranges (presented in in mm)

From: Preclinical investigations towards the first spacer gel application in prostate cancer treatment during particle therapy at HIT

  

Calculated

Session 1

Session 2

Session 3

#1

Reference

86.65

86.65

86.68

85.96

#2

Phantom-reference

86.65

86.61

86.69

85.97

#3

Empty vial

70.69

70.80

70.80

70.00

#4

Vial filled with water

38.51

38.71

38.72

38.02

#5

Vial filled with gel

-

[3] 38.31 ±0.30

[5] 38.27 ±0.29

[3] 37.63 ±0.29

#6

Vial filled with gel

-

-

-

[3] 37.68 ±0.51

 

(After 33 Gy irradiation)

    

#7

WEPL of spacer gel

-

1.0126 ±0.0111

1.0141 ±0.0101

1.082 ±0.0108

  1. Sessions 1–3 (columns) refer to three measurement series acquired within 10 weeks. The values in rows #1 to #6 are given in mmH2O. Rows #1 to #4 present the ranges acquired in reference measurements (2σ=0.2 mmH2O), row #5 (2σ uncertainty) reports the measured ranges of spacer gel, row #6 (2σ uncertainty) the measured range after the high dose irradiation (c.f., Section Insensitivity to high dose irradiation). In squared brackets [x] the number of performed measurements is stated. The bottom row (#7) presents calculated WEPL values.