Table 1 Dose differences (Measured – Calculated) in the centre of Ø1.5 cm and Ø4.0 cm tumours of the single anterior field irradiations and the standard deviations (SD) of the three discrete film measurements

From: A dosimetric phantom study of dose accuracy and build-up effects using IMRT and RapidArc in stereotactic irradiation of lung tumours

  

Chamber measurements

Film measurements, single field irradiations

  

Homogeneous Phantom

Heterogeneous Phantom, Perpendicular

Heterogeneous Phantom, Parallel

  

Measured - AAA

Measured - PBC

Measured - AAA

Measured - PBC

SD

Measured - AAA

Measured - PBC

SD

  

(cGy)

(cGy)

(cGy)

(cGy)

(cGy)

(cGy)

(cGy)

(cGy)

Ø1.5 cm

10 × 10 cm2

−1,3

−0,8

−0,9

−5,1

1,5

−1,8

−6,4

0,9

2.5 × 2.5 cm2

−0,5

0,7

−0,7

−6,0

1,8

0,3

−3,9

1,8

SWA 2 mm

5,7

6,6

5,8

−0,3

3,7

3,1

−1,6

1,8

SWA 4 mm

3,4

4,4

5,1

−0,9

2,0

3,9

−0,8

2,7

SWA 6 mm

2,0

3,0

2,8

−3,1

2,1

3,6

−1,0

2,5

SWA 10 mm

1,9

2,9

3,4

−2,5

1,1

3,1

−1,6

1,8

SWA 15 mm

1,4

2,4

1,5

−4,2

0,5

2,7

−1,9

2,0

Ø4.0 cm

10 × 10 cm2

−1,3

−0,8

−0,7

−4,6

0,3

−0,6

−4,4

1,4

5.0 × 5.0 cm2

−0,7

0,2

−0,1

−0,4

0,7

2,3

2,3

0,6

SWA 2 mm

1,4

2,4

1,5

1,7

1,4

−1,6

−1,1

0,9

SWA 4 mm

0,8

1,9

2,1

2,3

1,9

2,0

2,5

2,6

SWA 6 mm

0,8

1,9

1,5

1,6

1,0

0,6

1,1

1,7

SWA 10 mm

0,3

1,4

1,2

1,3

1,7

4,2

4,7

1,8

SWA 15 mm

0,5

1,6

1,1

1,3

1,7

4,0

4,5

1,1