Table 1 Deviation analysis - single film, lateral motion

From: Scanned carbon beam irradiation of moving films: comparison of measured and calculated response

    

ROIirradiated

ROItarget

Figure 2

A [mm]

φ [°]

T [s]

Mean [%]

SD

Mean [%]

SD

a)

0

  

-2.30

4.92

-1.52

2.06

b)

8

0

3

-2.98

16.92

-0.07

6.31

c)

8

0

4

-2.83

15.56

-0.98

9.16

d)

8

0

5

-2.24

18.53

0.73

9.34

e)

15

0

3

-1.85

10.78

0.66

6.19

f)

15

90

3

-0.89

11.53

0.73

9.50

g)

15

0

4

-3.50

12.06

-0.36

6.74

h)

15

90

4

-4.39

11.19

-1.23

5.17

i)

15

0

5

-5.40

14.99

-1.45

7.75

k)

15

180

5

-5.74

12.74

-2.50

7.61

l)

15

270

5

-6.46

16.09

-2.03

6.35

m)

20

0

5

-7.32

21.52

-2.92

14.49

n)

20

90

5

-5.03

20.64

-0.13

13.62

o)

20

0

7

-4.80

21.76

-0.34

15.08

p)

20

90

7

-9.84

68.98

-2.94

17.02

Mean

   

-4.52

19.52

-0.92

9.60

SD

   

2.39

14.74

1.30

3.86

Minimum

   

-9.84

10.78

-2.94

5.17

Maximum

   

-0.89

68.98

0.73

17.02

  1. Statistical analysis of relative deviations between calculated and experimental film response in two regions-of-interest (ROI). The film response distributions are shown in Figure 2 for the 14 different motion parameter combinations with amplitude A, period T, and phase φ at start of irradiation.