Skip to main content

Table 3 OSL detector dose measurements for cycle two of first patient for reduced MUs

From: A technique for pediatric total skin electron irradiation

OSL Detector Location AP + LPO + RPO Day 3 Dose (cGy) PA + LAO + RAO Day 4 Dose (cGy) Total Dose (cGy) % Diff from Prescription
Forehead 96.1 117.2 213.3 6.6
Posterior of Head 121.9 84 205.9 3.0
Umbilicus 79.1 117.3 196.3 18
Posterior of Umbilicus 125.6 72.6 198.2 09
Back of Right Hand 70.5 125.3 195.7 21