Table 1 OSL detector measured dose on 20 cm diameter cylindrical phantom following delivery of a prescribed 50 cGy

From: A technique for pediatric total skin electron irradiation

OSL Detector Location NanoDot 1 Dose (cGy) NanoDot 2 Dose (cGy) Average Dose (cGy) % Diff from Prescription
AP 54.3 53.3 53.8 7.6
Lt Lat 53.5 54.4 54.0 8.0
PA 53.9 53.1 53.5 7.0
Rt Lat 54.5 52.4 53.4 6.8
Average    53.7 7.3