Type of error
|
Explanation
|
Change in frequency and pulse width
|
Alarm is sounded
|
ICD reprogramming
|
Number of sessions detected by the programmer
|
Number of errors revealed by analysis of the ICD data logs
|
---|
Hard error
|
The ICDs stop generating pulses permanently.
|
Yes
|
No
|
Impossible
|
0
|
0
|
Soft error
| | | | |
13
|
29
|
Power-on reset
|
The ICDs present sudden complete failure and switch over to the safety backup mode.
|
Yes
|
Yes
|
Necessary
|
8
|
8
|
Partial electrical reset
|
The ICDs show no change in mode of pacing.
|
No
|
Yes
|
Unnecessary
|
5
|
7
|
Minor error
|
The ICDs show no change in mode of pacing.
|
No
|
No
|
Unnecessary
|
0
|
14
|
- Abbreviation: ICD = implantable cardioverter defibrillator