Power | 0dB, Mechanical index = 1.6: As Low As Reasonably Achievable to minimize bio-effects. |
Time Gain Compensation (TGC) | Adjusted to compensate the effect of attenuation in the subcutaneous tissue at greater depths in order to produce images of uniform brightness. |
Dynamic Range | 66 dB permits best differentiation between subtle changes in echo intensities of the skin region. |
Persistence | 2 used to provide optimum smoothening of images with minimum movement induced artifacts. |
Post-processing | Options range from low (0) to high (3) contrast. It was at 2 to obtain optimal level of contrast. |
Overall Gain | Set at 2dB to demonstrate tissues with appropriate brightness. |
Edge | +2 to emphasize the boundaries between tissues. |
Delta | Δ2 for high contrast resolution. |
Spatial Compounding | A form of frame averaging. SC1 selected to maximize frame rate. |
Frame Rate | Decreases with increasing the scan depth, the number of focal zones and image line density (resolution). Varies during the scan but it should not be below 10 Hz (fps). |
Focal Zone | Single focal zone placed at the level of the skin to optimize lateral resolution. |