Skip to main content

Table 4 Volume error (VE), centroid error (CE) and JAC index for left lung.

From: Validation of an elastic registration technique to estimate anatomical lung modification in Non-Small-Cell Lung Cancer Tomotherapy

  

kVCT/1st MVCT

kVCT/2nd MVCT

kVCT/3rd MVCT

patient

registration

VE %

C E

JAC

VE %

C E

JAC

VE %

C E

JAC

1.

rigid

24.85

7.35

0.72

-17.87

9.76

0.76

45.14

7.86

0.54

 

elastic

1.97

0.40

0.95

-0.81

0.14

0.96

3.10

0.84

0.92

2.

rigid

-5.14

1.22

0.90

-5.55

2.05

0.89

2.64

0.93

0.89

 

elastic

0.25

0.39

0.94

1.49

0.26

0.95

1.11

0.76

0.92

3.

rigid

-1.97

2.15

0.80

-4.53

2.58

0.87

-17.38

5.15

0.81

 

elastic

1.99

1.94

0.89

0.01

2.11

0.90

-6.82

3.23

0.88

4.

rigid

-0.80

2.00

0.91

-5.42

2.14

0.90

-10.69

4.06

0.84

 

elastic

0.80

0.06

0.95

-0.59

0.37

0.96

-2.08

0.56

0.95

5.

rigid

2.35

0.89

0.92

4.60

0.96

0.93

4.26

0.99

0.90

 

elastic

2.23

0.16

0.95

0.49

0.09

0.96

2.05

1.52

0.94