Table 1 Statistical analysis of high dose point in the PTV measured with ion-chamber: [(measured dose)-(calculated dose)]/calculated dose, with associated confidence limit

From: Statistical analysis of IMRT dosimetry quality assurance measurements for local delivery guideline

Treatment site

Location

Mean

Standard deviation (σ)

Maximum

Minimum

Prostate

I/C1 or near I/C

-0.004

(-0.4%)

0.012

(1.2%)

0.025

(2.5%)

-0.046

(-4.6%)

Brain

I/C or near I/C

-0.006

(-0.6%)

0.015

(1.5%)

0.039

(3.9%)

-0.041

(-4.1%)

Head and Neck

I/C or near I/C

-0.016

(-1.6%)

0.012

(1.2%)

0.027

(2.7%)

-0.046

(-4.6%)

Overall combined

-0.010

(-1.0%)

0.015

(1.5%)

0.039

(3.9%)

-0.046

(-4.6%)

Confidence limit = (|mean|+1.96σ)

0.038 (3.8%)

  1. Note: 1. if doses at the isocenter were not fairly uniform, then the ion-chamber was relocated in the region of uniform dose near isocenter. I/C stands for isocenter.