Figure 1 | Radiation Oncology

Figure 1

From: Two years experience with quality assurance protocol for patient related Rapid Arc treatment plan verification using a two dimensional ionization chamber array

Figure 1

a) Example of measured 2D dose distribution of a head and neck case. The MLC positions were changed in the region of the spinal cord to got higher dose to the spinal cord which could not be clinically tolerated. b) Passing rate against the maximum dose to the spinal cord of the same head and neck case. 4 Rapid Arc treatment plans were generated by changing the MLC positions. The changed Rapid Arc treatment plans were compared to the original Rapid Arc treatment plan. The original Rapid Arc treatment plan showed a passing rate of 99.3%. The lines indicate the in our clinic tolerated limits: 45 Gy maximum dose to the spinal cord and 99% passing rate.