Table 4 Gamma analysis distribution for the different treatment plans in inhomogeneous thorax phantom for different acceptance criteria, for EDR2 films.

From: Dosimetric accuracy of tomotherapy dose calculation in thorax lesions

  EDR2 film
  3%-3 mm 4%-4 mm 5%-4 mm
  0.7 1 1.5 0.7 1 1.5 0.7 1 1.5
Med. 1(film 1) 91 96 99 95 98 100 96 98 100
Med. 1 (film2) 92 96 98 94 97 100 95 97 100
Med. 2(film 1) 89 91 93 90 93 95 91 93 95
Med. 2 (film2) 81 89 94 84 91 94 87 93 94
Single Met. (film 1) 88 96 97 92 96 98 94 97 99
Single Met. (film 2) 88 93 97 91 95 99 93 96 99
Multipla Met. (film 1) 93 95 98 93 96 98 96 97 98
Multipla Met. (film 2) 85 93 99 89 96 100 92 97 100
  1. Percentage of points with γ value ≤ 0.7, 1 and 1.5 was reported, separately for film 1 and film2.