Skip to main content

Table 3 Statistical analysis for the ionization chamber array measurements.

From: Dosimetric precision of an ion beam tracking system

IC position

Mean

SD

abs. mean

abs SD

minimum

maximum

target volume

0.3

1.5

1.2

0.9

-2.7

3.7

left penumbra

-13.3

2.3

13.3

2.3

-15.8

-10.5

right penumbra

7.8

0.1

7.8

0.1

7.8

7.9

distal edge

-1.5

5.1

4.1

3.1

-8.9

3.3

position A

0.8

1.6

1.4

1.0

-2.5

3.7

position B

-0.2

1.3

1.0

0.8

-2.7

2.3

  1. All values denote the relative dosimetric deviation between motion compensated and the stationary reference measurement in %. Position A refers to chamber positions indicated by solid lines, position B to chamber positions indicated by thin lines in fig. 1c.