Skip to main content

Table 2 Mean, standard deviation and ranges of volumetric coverage and percent dose delivered to selected target and normal tissues as defined in Table 1 for three dose calculation algorithms during whole breast tangent radiation therapy.

From: The impact of dose calculation algorithms on partial and whole breast radiation treatment plans

 

DEV

[%]

PTV

[%]

IPS-BR

[%]

IPS-LUNG

[%]

HEART

[%]

CON-LUNG

[%]

CON-BR

[%]

PBC

97.4 (4.3)

80.3 (8.7)

67.1 (5.9)

15.1 (6.3)

12.8 (15.9)

1.1 (0.9)

18.6 (29.7)

 

87.4-100.0

60.7-93.7

60.8-76.6

7.1-26.1

1.6-47.0

0.0-2.6

1.4-91.6

AAA

92.5 (8.5)

73.4 (9.3)

56.6 (7.9)

21.2 (7.6)

12.6 (16.0)

1.0 (0.6)

23.7 (25.0)

 

72.0-100.0

60.6-91.3

41.0-69.1

10.9-33.6

1.3-47.0

0.8-2.2

3.0-101.1

MC

94.4 (5.5)

75.9 (13.7)

55.3 (9.4)

19.9 (6.2)

12.4 (15.6)

1.1 (0.5)

19.3 (26.8)

 

83.9-100.0

60.5-98.4

42.6-69.3

10.6-31.1

1.3-44.3

0.5-2.0

5.6-94.1