Skip to main content

Table 2 Agreement between the 2D dose distributions as measured with film and calculated with MC and fsPB for the 3%/3 mm and 2%/2 mm γ acceptance criteria; the acceptance (in %) is listed for the whole region (all), the tumour region (tumour) and the region outside the tumour (lung) and for all tumour sizes.

From: Monte Carlo vs. Pencil Beam based optimization of stereotactic lung IMRT

 

MC 3%/3 mm

fsPB 3%/3 mm

MC 2%/2 mm

fsPB 2%/2 mm

Tumour diameter [cm]

all

tumour

lung

all

tumour

lung

all

tumour

lung

all

taret

lung

 

6 MV

2.7

99.6

99.6

99.6

71.8

67.4

72.2

97.3

97.1

97.3

43.3

43.1

43.3

4.2

99.3

99.0

99.4

49.4

31.2

53.1

96.0

92.0

96.8

28.5

23.7

29.5

5.0

97.0

97.2

97.0

46.6

31.3

51.4

89.4

94.4

87.8

30.5

21.2

33.4

 

15 MV

2.7

99.9

99.6

99.9

81.4

60.6

83.0

97.4

96.8

97.4

52.2

41.4

53.0

4.2

97.8

99.6

97.4

51.2

42.4

53.0

84.0

96.7

81.4

25.3

27.7

24.8

5.0

98.3

98.9

98.1

45.4

37.7

47.8

89.8

93.9

88.5

26.6

22.8

27.8