From: Monte Carlo vs. Pencil Beam based optimization of stereotactic lung IMRT
6 MV | 15 MV | ||||||||
---|---|---|---|---|---|---|---|---|---|
Quantity type | Prescr. | PB | 4D | MC | 4D | PB | 4D | MC | 4D |
model 1, one phase static CT | |||||||||
target* poisson EUD | 55 Gy | 54.25 (57.94) | 58.86 | 53.3 | 57.35 | 55.32 (52.77) | 55.47 | 52.09 | 58.82 |
target* rms overdosage | 2 Gy | 1.99 (4.17) | 4.48 | 1.53 | 3.39 | 2.02 (0.66) | 1.95 | 1.97 | 5.83 |
lung R, mean dose | 9 Gy | 8.39 (10.58) | 10.81 | 9.1 | 9.32 | 8.46 (10.31) | 10.47 | 9 | 9.26 |
lung R, mean damage | 21% | 20.49 (24.48) | 24.93 | 21.68 | 22.05 | 20.71 (23.9) | 24.21 | 21.17 | 21.6 |
skin, rms overdosage | 0.24 Gy | 0.23 (0.61) | 0.59 | 0.24 | 0.24 | 0.24 (0.52) | 0.52 | 0.24 | 0.24 |
model 2, minimum density overwrite one phase static CT | |||||||||
target* poisson EUD | 55 Gy | 56.47 (57.23) | 57.53 | 54.98 | 56.2 | 55.95 (53.68) | 54.37 | 52.47 | 56.02 |
target* rms overdosage | 2 Gy | 2.05 (3.23) | 2.83 | 2.02 | 1.8 | 2.02 (0.96) | 0.6 | 2.07 | 2.6 |
lung R, mean dose | 9 Gy | 8.13 (9.63) | 9.84 | 8.91 | 9.13 | 8.34 (9.96) | 10.06 | 8.71 | 8.88 |
lung R, mean damage | 21% | 19.9 (22.57) | 23.03 | 21.25 | 21.66 | 20.43 (23.14) | 23.38 | 20.59 | 20.88 |
skin, rms overdosage | 0.24 Gy | 0.23 (0.48) | 0.48 | 0.24 | 0.23 | 0.24 (0.48) | 0.48 | 0.24 | 0.23 |
model 3, average CT | |||||||||
target* poisson EUD | 55 Gy | 55.03 (57.53) | 57.46 | 52.63 | 55.88 | 55.32 (52.57) | 55.01 | 51.85 | 58.49 |
target* rms overdosage | 2 Gy | 1.99 (3.71) | 3.09 | 1.12 | 1.66 | 2 (0.52) | 1.25 | 2.01 | 4.63 |
lung R, mean dose | 9 Gy | 8.43 (10.41) | 10.64 | 9.12 | 9.33 | 8.49 (10.3) | 10.49 | 9.04 | 9.3 |
lung R, mean damage | 21% | 20.56 (24.14) | 24.61 | 21.74 | 22.13 | 20.72 (23.85) | 24.23 | 21.21 | 21.66 |
skin, rms overdosage | 0.24 Gy | 0.24 (0.61) | 0.6 | 0.24 | 0.23 | 0.24 (0.53) | 0.53 | 0.24 | 0.23 |