Skip to main content

Table 6 EDW: average (± SD) percentage difference of Dose/MU between GLAaS and Reference (Eclipse or ion chamber), and range over 14 analysed EDW fields.

From: Testing the portal imager GLAaS algorithm for machine quality assurance

 

GLAaS vs Eclipse [%]

GLAaS vs IonCh. [%]

Eclipse vs IonCh. [%]

Dose/MU:

-1.1 ± 1.6 [-4.8, +1.2]

0.4 ± 0.7 [-0.9, +1.4]

1.6 ± 1.3 [+0.1, +4.2]