Skip to main content

Table 2 Population systematic (Σ) and random (σ) errors of selected contemporary series and correlation with probability of target volume coverage

From: Assessment of three-dimensional set-up errors in conventional head and neck radiotherapy using electronic portal imaging device

Series

Σ

σ

Displacements or errors

Hess [10]

Not reported

Not reported

3 mm for 50% coverage

   

9 mm for 95% coverage

Bentel [11]

Not reported

Not reported

5–10 mm (87–90% with 5 mm margin)

Gibeau [12]

1 – 2.2

0.7 – 2.3

4.5–5.5 mm for 90%probability of target coverage

De Boer [13]

1.5 – 2.0

1.5 – 2.0

Probability values not specified

Humphrey [14]

0.02 – 0.9

0.4 – 0.7

3 mm for 95% of the errors.

   

5 mm for 99% of errors

Zhang [15]

1.5 – 3.2

1.1 – 2.9

5.5 mm for 90% probability of target coverage

Suzuki [16]

0.7 – 1.3

0.7 – 1.6

5 mm margin for PTV and 3 mm for PRV

   

Probability values not specified

Present Study

0.96 – 1.2

1.94 – 2.48

93% displacements within 5 mm

   

<5 mm CTV-PTV margin in all directions