From: Phantom investigation of 3D motion-dependent volume aliasing during CT simulation for radiation therapy planning

Figure 4

Fast- and slow-scan distortion of the 10- and 31.5-mm diameter targets as a function of four (4) initial motion phases and three (3) motion amplitudes. The top row of images ("a" and "b") is associated with the 10 mm target, while the bottom row ("c" and "d") with the 31.5 mm target. The columns of structures labeled "STATIC" are surrogate representations of the respective 10- and 31.5-mm diameter targets. Image sets "a" and "c" are reconstructions from single-slice fast techniques, while "b" and "d" are from single-slice slow scan techniques. The motion amplitudes presented on the figures are for the S-I direction and are synchronized with constant ± 2 and ± 5 mm displacements in the R-L and A-P directions, respectively.