Skip to main content

Table 1 MCNP derived results for BLOKX model.

From: Novel low-kVp beamlet system for choroidal melanoma

 

Location

Dose (Gy)

Uncertainty (%)

Tumor 90° from macula

Base

101.8

0.32

 

Apex

85.0

0.35

 

Macula

1.5

2.64

 

Optic Nerve

1.6

2.54

 

Lens

3.3

1.75

Tumor 60° from macula

Base

87.1

0.36

 

Apex

85.0

0.36

 

Macula

3.1

1.9

 

Optic Nerve

1.2

3.06

 

Lens

2.4

2.19

Tumor 45° from macula

Base

85.0

0.4

 

Apex

93.1

0.38

 

Macula

51.0

0.52

 

Optic Nerve

1.8

2.72

 

Lens

3.6

1.9

Tumor 45°R* from macula

Base

85.0

0.46

 

Apex

112.0

0.4

 

Macula

4.0

2.1

 

Optic Nerve

2.9

2.45

 

Lens

7.1

1.57

Tumor 30° from macula

Base

85.0

0.42

 

Apex

99.1

0.39

 

Macula

70.4

0.46

 

Optic Nerve

3.0

2.23

 

Lens

4.7

1.75

Tumor 270° from macula

Base

102.8

0.31

 

Apex

85.0

0.35

 

Macula

3.2

1.8

 

Optic Nerve

2.9

1.88

 

Lens

1.8

2.36