4DCT and CBCT based PTV margin in Stereotactic Body Radiotherapy(SBRT) of non-small cell lung tumor adhered to chest wall or diaphragm

Abstract

Background

Large tumor motion often leads to larger treatment volumes, especially the lung tumor located in lower lobe and adhered to chest wall or diaphragm. The purpose of this work is to investigate the impacts of planning target volume (PTV) margin on Stereotactic Body Radiotherapy (SBRT) in non-small cell lung cancer (NSCLC).

Methods

Subjects include 20 patients with the lung tumor located in lower lobe and adhered to chest wall or diaphragm who underwent SBRT. Four-dimensional computed tomography (4DCT) were acquired at simulation to evaluate the tumor intra-fractional centroid and boundary changes, and Cone-beam Computer Tomography (CBCT) were acquired during each treatment to evaluate the tumor inter-fractional set-up displacement. The margin to compensate for tumor variations uncertainties was calculated with various margin calculated recipes published in the exiting literatures.

Results

The means (±standard deviation) of tumor centroid changes were 0.16 (±0.13) cm, 0.22 (±0.15) cm, and 1.37 (±0.81) cm in RL, AP, and SI directions, respectively. The means (±standard deviation) of tumor edge changes were 0.21 (±0.18) cm, 0.50 (±0.23) cm, and 0.19 (±0.44) cm in RL, AP, and SI directions, respectively. The means (±standard deviation) of tumor set-up displacement were 0.03 (±0.24) cm, 0.02 (±0.26) cm, and 0.02 (±0.43) cm in RL, AP, and SI directions, respectively. The PTV margin to compensate for lung cancer tumor variations uncertainties were 0.88, 0.98 and 2.68 cm in RL, AP and SI directions, which were maximal among all margin recipes.

Conclusions

4DCT and CBCT imaging are appropriate to account for the tumor intra-fractional centroid, boundary variations and inter-fractional set-up displacement. The PTV margin to compensate for lung cancer tumor variations uncertainties can be obtained. Our results show that a conventional 1.0 cm margin in the SI plane dose not suffice to compensate the geometrical variety of the tumor located in lower lobe and adhered to chest wall and diaphragm.

Background

Lung cancer is one of the leading causes of death in cancer patients in China. According to the literature, 3-year local control rate of the lung cancer is only 66% in conventional conformal radiotherapy [1]. The stereotactic body radiation therapy (SBRT) is a new technology which can effectively improve the treatment effect of lung cancer. A meta-analysis of 34 lung cancer SBRT studies including 2578 patients reported that 3-year local control rate for tumors measuring greater than 3 cm was 87% [2]. SBRT delivers high dose in fewer fractions (3–10 fractions) than conventional radiotherapy which may increase toxicity in adjacent organs at risk (OAR). It is therefore critical to minimize the irradiated volume of healthy tissue when delivering the prescribed dose to tumor in SBRT.

However, a variety of geometrical uncertainty such as respiratory motion [3, 4], baseline variation [58] and set-up error [9], limit the precision of radiation therapy (RT) for lung cancer. According to ICRU report 62#, internal margin (IM) and set-up margin (SM) should be included in the PTV to compensate geometrical uncertainties including tumor centroid movement [10], tumor boundary [11, 12] and set-up displacement, especially in SBRT. However, the tumor boundary displacement merely discussed about margin calculation previously. It would be assessed in present study.

As a standard modality to compensate tumor motion, four-dimensional CT (4DCT) has been generally applied to the calculation of the extent margin of tumor motion [13], and is also incorporated into treatment planning. When only areas with tumor appearance are included in the target volume definition, there is a significant reduction of the mean volume by the use of 4DCT [14]. Generally, the need of contour gross tumor volume (GTV) in up to ten phases of respiration is a time-consuming procedure for the routine clinical use of 4DCT scans. To reduce the workload of multiple contour delineations in 4DCT, maximum intensity projection (MIP) and averaged intensity (AI) of 4DCT image are reconstructed in order to be used in the treatment planning, MIP provide a 3D CT scan whose vowels’ values are the greatest vowels’ intensity values throughout the 4D CT dataset. AI provides a 3D CT scan whose vowels’ values are the arithmetic mean of the 4D CT scans [15]. However, the tumor contoured with the MIP or AI series cannot totally represent the true tumor size, location and shape, especially adhered to chest wall or diaphragm [15]. To correctly assess the tumor motion and tumor shape, ten phase of the 4DCT were contoured in our study.

Tumor localization is also a significant issue in NSCLC SBRT, the major focus has been the development of image-guidance system. Several image guidance techniques such as electronic portal imaging device (EPID) [16] and CBCT [17], have been adapted to narrow the set-up margin. Among them, three-dimensional (3D) matching, coupled with CBCT provides results that have greater accuracy than two-dimensional matching, coupled with EPID [18]. Generally, with the repeating CBCT imaging, intra-fractional and inter-fractional set-up displacement can be assessed for the lung cancer patients’ set-up displacement using KV X-ray,such as fluoroscopy, CBCT and four dimensional CBCT (4D-CBCT) [1921]. CBCT imaging is sufficient to account for the inter-fractional tumor variation and setup variation based on the bone structure, which was adopted in our study. Furthermore, size, patients characteristics are also significant factors to influent patients’ set-up margin [22], but these factors are negligible in the set-up margin.

The present study would evaluate changes in tumor motion magnitude and set-up error by 4DCT at planning and CBCT at treatment, and calculated the clinical target volume (CTV)-planning target volume (PTV) margins to compensate for these changes, which may minimize the risk of complications and tumor failures. The aim of the present study is to decide whether conventional 1.0 cm margin in the SI plane and 0.5 cm margin in the other plane suffice to compensate the geometrical variety of the tumor adhere to chest wall and diaphragm.

Methods

Patient and tumor characteristics

Between February 2013 and June 2015, 20 patients were recruited to this study to assess PTV margin. Table 1 summarizes the patient and tumor’s general information and detailed characteristics.

Table 1 Patients and tumor generally characteristics of 20 patients in this study

The studies were approved by the Institutional Committee for Clinical Research and the Local Ethics Committee in The First Affiliated Hospital, Xi’an Jiaotong University of medical college. Eligible patients were given written informed consent.

4DCT simulation and image acquisition

During the CT simulation, all patients were immobilized using thermoplastic mask in the supine position with arms raised above the head. Then, each patient received a helical treatment planning three-dimensional computed tomography (TP-3DCT) and an additional 4DCT simulation under free breathing condition on a 16-slice CT scanner (Big bore, Philips Medical Systems, Cleveland, OH). For TP-3DCT, each scan’s rotation time took about 1 s to acquire image dates. The other parameters were as follows: pitch 0.85, 120KV and 400mAs. During 4DCT scanning, the respiratory signal was acquired by a pressure sensor (“pneumonia bellows”). The pressure in the bellows decreases as the patient inhales, with the opposite happening as the patient exhales. Then, the pressure change was measured by a remote pressure sensor which provided a direct-current voltage to an analog-to-digital converter. 4DCT images were reconstructed and then sorted into 10 respiratory phases, with 0% representing end-inhalation and 50% representing end-exhalation. Prospectively phase-binned image sets was been used to extract displacement binned 4DCT. Both the TP-3DCT and 4DCT images were reconstructed with a thickness of 3 mm and then transferred to a Pinnacle V. 9.0y treatment planning system (Philips Medical Systems, Fitchburg, USA).

Manual contouring and treatment planning

For target volume definition, the 4DCT phases display the extreme of tumor positions for each spatial direction. After the matching of CT-cubes between 4DCT phases and 3DCT phase in the Philips treatment planning system, GTV contours were manually delineated on the 10 phases of the 4DCT scan using the lung window setting by the same physician to make the inter-observer lowest as soon as possible (L = −600Hu, W = 1600Hu [20]), and CTV4D was defined as GTV4D plus an isotropic margin of 0.6 cm for adenoid carcinoma and the one of 0.8 cm for squamous carcinoma. The combined volume of each CTVs in 10 phases was defined as internal CTV4D (ITV). The modification to form the PTV4D was performed by adding an isotropic margin of 3 mm to ITV to account for set-up inaccuracies in all direction (Fig. 1). The PTV4D contours on 4DCT phases were projected to the contour named PTV on the TP-3DCT.

Fig. 1
figure 1

Phase 0% of four-dimensional computed tomography (4DCT) image for an protocol NSCLC patient. (a) transverse (b) sagittal (c) coronal; Gross tumor volume (GTV) of phase 0% in Red; Clinical target volume (CTV) of phase 0% in Green; Internal clinical target volume (ITV) in blue; Planning target volume (PTV) in Yellow

For treatment planning, every patient received an individual treatment plan and the plan calculation were performed on 3DCT, 10 patients were treated with the IMRT technology, and the others with conformal arc technology. All patients were prescribed total doses of 48Gy in 4 fractions [16, 23, 24] with the 95% isodose covering the PTV.

Tumor boundary change definition

Figure 2 shows tumor boundary change definition. Ten beams with gantry angle 0° and 90° were used and conformed to GTV of each phase with centroid of corresponding GTV for each patient. The values of jaw location in the RL, and SI directions with beam of angle 0° were collected. Then the same values with beam of angle 90° were also collected. The variation of these values represent the boundary displacement in the RL, AP and SI direction.

Fig. 2
figure 2

Phase 0% of four-dimensional computed tomography (4DCT) image for an protocol NSCLC patient. (a) coronal (b) sagittal; Gross tumor volume (GTV) of phase 0 % (GTV0%) in Red; Gross target volume (GTV) of phase 50 % (GTV50%) in Green

CBCT simulation and image acquisition

The on-board imager integrated in a Synergy medical linear accelerator (Elekta, Stockholm, Sweden) was adopted to acquire CBCT images. Before acquiring the image data, each patient was immobilized in the same position as that used in 4DCT scan. A 120 kV x-ray tube voltage and 400mAs current were used for generating middle-resolution images. The scan angle was from −178° to 178°. All CBCT images were reconstructed with a thickness of 3 mm. After each patient were initial positioned, a CBCT was acquired and registered to the planning TP-3DCT based on the location of vertebral bodies (bony landmarks) (Fig. 3). The target position error was corrected by shifting the treatment couch. If the position error was ≤2 mm in all three direction, the patient was treated without shifting couch; else couch was shifted and a second CBCT was acquired to measure the residual error until a residual error was ≤2 mm [19]. At the end of treatment, a CBCT scan was finally performed to assess the intra-fractional set-up displacement.

Fig. 3
figure 3

Precorrection cone-beam CT (CBCT) image for an protocol NSCLC patient in the thermoplastic mask. TP-3DCT GTV in Yellow; Precorrection GTV in Red

Respiration motion analysis

Centroid position and boundary of the respective CTV4D were calculated and compared. The centroid shift of respective CTV4D was represented as tumor position variation, and the tumor boundary change of respective CTV4D was represented as tumor shape variation during respiration. The 3D spatial motion vector of the individually CTV4D centroid and boundary variation were evaluated according to the recipe as follows [25]

$$ V=\sqrt{RL^2+{AP}^2+{SI}^2}. $$

Margin calculation

Existing literature has proposed approaches to calculate the PTV margin based on systematic and random uncertainties (see Table 2) [8, 2628], which were applied for the PTV margin calculation in our study. Among these approaches, the recipes approached by Stroom and van Herk are widely adopted in current clinical treatment. In all recipes, ∑ represents the standard deviation (SD) of systematic errors and σ is the root-mean square of random errors. In this study, the systematic errors include the standard deviat55ion (SD) of the centroid movement denoted as ∑ centroid , the tumor boundary changes denoted as ∑ boundary and set-up displacement denoted as ∑ setup of each patient. As each patient had one 4DCT scan, the random errors of the centroid and boundary changes were zero, and the total random errors equal to set-up random errors. As a result, the total SD of systematic errors ∑ total and random errors were evaluated with the recipe: \( {\sum}_{total}=\sqrt{\sum_{centroid}^2+{\sum}_{boundary}^2+{\sum}_{setup}^2} \), σ total  = σ setup .

Table 2 Summary of various published recommendations for margins around target volumes (CTV)

Data analysis

SPSS 19.0 was used for statistical analysis. Differences were considered significant for P <0.05. For test of correlation, it is Person’s rank test which was used.

Results

Tumor centroid and boundary changes

Figures 4 and 5 show the tumor boundary and centroid changes for 20 patients. A positive value in the right-left (RL), anterior-posterior (AP), and superior-inferior (SI) directions represents a shift in the right, anterior, and inferior directions, respectively. We did not find any statistically significant correlation between tumor centroid displacement or boundary changes and patients’ age. The mean 3D centroid change value was 1.55 ± 0.46 cm, with a maximum at 2.33 cm. Tumor centroid had a greater mean changes in SI direction than in other direction (SI vs RL: t = 0.587,P < 0.05; SI vs AP: t = 0.547, P < 0.05). The mean 3D boundary change value was 0.70 ± 0.33 cm, with a maximum at 1.40 cm. Tumor boundary motion had a greater mean displacement in AP direction than in other directions, but it is not statistically significant, which is similar to the one observed by Gauthier et al. [12]. In all direction, the average boundary changes directions were toward to the centroid changes directions, which means the tumor shrank as expected.

Fig. 4
figure 4

Boundary errors in RL,AP,and SI directions and 3D vector for 20 patients 

Fig. 5
figure 5

Centroid position errors in RL,AP,and SI directions and 3D vector for 20 patients

Tumor set-up errors

Figure 6 shows tumor set-up errors in RL, AP, and SI directions for 80 pre-correction CBCT fractions. A positive value in the right-left (RL), anterior-posterior (AP), and superior-inferior (SI) directions represents a shift in the right, anterior, and inferior directions, respectively. The maximum set-up error were as large as 0.66 cm in RL direction, 0.54 cm in AP direction, and 1.16 cm in SI direction. The average setup errors are 0.13 ± 3.15 cm,0.37 ± 2.76 cm and 0.20 ± 5.73 cm in RL, AP and SI directions, respectively. From the distribution, we could calculated the probability of having set-up errors lager than 5 mm (which is the margin used for set-up margin) and found it to be 15.0% in RL, 6.3% in AP, and 38.8% in SI directions, respectively. The mean set-up errors had a greater value in SI direction than in other directions. (SI vs AP: r = −0.609, p <0.05; SI vs RL: r = −0.480, p > 0.05).

Fig. 6
figure 6

Set-up errors in in RL, AP, and SI directions and 3D vector for 80 fractions. Data were derived based on bony landmark alignments

PTV margin

Table 3 shows the SDs of systematic (∑) and random (σ) errors. One patient had one 4DCT scan at tumor CT simulation, and therefore the random errors were zero for tumor centroid and boundary displacement. Based on the SDs of set-up systematic and random errors in previous table, the pre-correction set-up margin was calculated and showed in Table 4. Among all recipes, the maximum were 0.67, 0.69 and 1.21 cm in RL, AP and SI directions, which were calculated with the recipe by van Herk et al. To compensate all geometrical uncertain motioned above, the total margin were calculated and showed in Table 4. Among all recipes, the maximum were 0.88, 0.98 and 2.68 cm in RL, AP and SI directions, which were also calculated with the recipe by van Herk et al.

Table 3 The Standard deviation of systematic (∑) and random (σ) errors
Table 4 CTV margin changes of NSCLC patients

Discussion

It is reported [29] that the tumor movement in lung is not related to patient height, weight, cancer stage and lung function, and therefore these issues were not included in this study. We evaluated changes in tumor motion magnitude and set-up error by 4DCT scan at planning and CBCT scan at treatment and calculated the margins to compensate for these changes [16, 17]. To our knowledge, this is the first study to report the evaluation of tumor centroid, boundary changes and set-up displacement, which are account into the margins calculation.

We admit that at the time we performed this study, we did not perform Post-correction and post-treatment CBCT to verify that set-up displacement and tumor locatioin had not changed. However, [30, 31] existing literatures have shown that post-correction margins based on bony structure are negligible to account into PTV margin, which are less than 2 mm in three directions [24]. Other literature has shown that inter-fractional tumor position and breathing motion based on bony structure during treatment are stable. Therefore, the post-treatment margins are also negligible to account into PTV margin. In addition, we did perform post-correction and post-treatment CBCT on other patients and found that the tumor with post-correction and post-treatment displacement changed is negligible, and therefore the data have not yet been published.

The variation in the centroid changes between ITV based from 4DCT and CTV based from 3DCT was not significant [32]. Therefore, the intra-fractional tumor motion on 4DCT different phase can represent the one on the TP-3DCT. Intra-fractional tumor motion in present study was divided into tumor centroid and boundary displacement under free-breathing conditions. Centroid and boundary variation were collected starting from 0% phase to ending at a phase of maximum variation, which was 50% phase mostly. So the variation of centroid and boundary were collected starting from inferior direction to ending at superior direction. In my study, A positive value in superior-inferior (SI) directions represents a shift in the inferior directions. So the values for centroid an boundary in the graphs are negative values. Mori, S et al. [33] quantified the magnitude of intra-fraction lung tumor motion under free-breathing conditions using 4DCT and reported the mean (±standard deviation) tumor centroid motion were 0.19 (±0.16), 0.40 (±0.23), and 1.03 (±0.71)mm in the RL, AP, and SI directions respectively, which was similar with the present study result. Only the motion in SI direction was less than present study. We presume that only a few lung tumors were located in the lower lobe on previous study, resulting in their small centroid changes in SI direction.

Tumor boundary motion, which was merely discussed on previous PTV margin study, were quantified in present study and found that the maximum were 0.57, 0.99, and 0.97 cm in RL, AP and SI direction respectively. We could calculated the probability of having boundary motion lager than 5 mm and found it to be 10.0% in RL, 40.0% in AP, and 7.5% in SI directions, respectively. We presume that the tumor located in lower lobe patients was adhered to the diaphragm, resulting in their significant boundary changes. We believe that the boundary changes in intra-fractional variation is significant, especially in SBRT treatment.

Set-up error in the present study was defined as the shift from TP-3DCT to CBCT. This error includes the inter-fractional set-up error based on bony structure. The standard deviation of set-up systematic errors (∑) are 0.24, 0.26 and 0.43 cm in RL, AP and SI directions. The result was similar with the one observed by Worm, E.S. et al., which was 0.23, 0.16 and 0.45 cm [31]. The standard deviation of set-up random errors (∑) are 0.10, 0.08 and 0.20 cm in RL, AP and SI directions. The result was also similar with the one observed by Li, W ea al, which was 0.12, 0.13 and 0.17 cm with Performance Status 0 patients stratified by Eastern Cooperative Oncology Group performance [24]. The average setup errors are 0.13 ± 3.15 cm,0.37 ± 2.76 cm and 0.20 ± 5.73 cm in RL, AP and SI directions, respectively. And the maximum set-up margin in the present study was 0.67, 0.69 and 1.21 cm in RL, AP and SI directions calculated with van Herk recipe. These result was similar with the study by Wang L et al. [23], which assessed set-up margin using three-dimensional CBCT and report it to be 0.9–1.0 cm range. However, the study by Ueda Y et al. [16] assessed set-up margin by shift from MTP-4DCT at planning to MTP-cine at the treatment and reported it to be 0.52 cm in SI direction in 98% session’s calculated Stroom recipe (Table 2). The margin in study above is less than that in present study. We presume that their set-up, which used Body Fix double-vacuum system, resulted in their smaller set-up error.

The PTV margin also depends on the patient positioning techniques. Corradetti MN et al. [34] Stated that margins were 0.49 cm, 0.85 cm and 0.77 cm in RL, AP, and SI direction calculated by the van Herk recipe, respectively, which is smaller than the present result. We presume that their set-up, which used the full-body vacuum cushion system, resulted in their smaller inter-fractional error. Many other studies on SBRT have adopted 5 mm as the PTV margin using various advanced immobilized devices, such as Microtron, BodyFix. These devices can obviously reduce the PTV margin in the SBRT treatment for NSCLC. However, these advanced devices cause more time in the patients’ set-up procedure and bring more variations in the time spent on the treatment couch [17].

A total PTV margin combining tumor centroid, boundary changes and set-up displacement were 0.88, 0.98 and 2.68 cm in RL, AP and SI directions, respectively, which were maximum among all margin recipes. According to the comments of Table 5, the statistical assumption with van Herk’s recipe is minimum absorbed dose to CTV is 95% for 90% of patients. It’s requirement about minimum absorbed dose is high in realistic treatment plans, so the bigger margin size are needed to meet the requirement. This appoint is proved in my study, which shows that among all recipes, the maximum margin were calculated with the recipe by van Herk et al. And the requirement of Parker’s recipe about minimum absorbed dose is the same as one of van Herk’s, but the probability lever is not specified. And the results in my study shows that the margin calculated with Parker’s recipe is less than one with van Herk’s recipe.

Table 5 Summary of various published recommendations for margins around target volumes

Many studies stated that PTV margin was 0.5 cm in the axial plane and 1.0 cm in the cranial-caudal plane in all directions in the study on SBRT for NSCLC [20, 3537]. The results were less than the one observed in SI direction of the present study. In conclusion, a 0.5 cm margin in the axial plane suffices to compensate the geometrical variety; nevertheless, a 1.0 cm margin in the SI plane is dose not compensated the geometrical variety in the present study, which the tumor located in lower lobe is adhered to chest wall and diaphragm.

Conclusions

4DCT imaging is appropriate to account for the intra-fractional tumor centroid position, and boundary variations, and CBCT imaging is appropriate to account for the inter-fractional tumor set-up variation. We observed that the large tumor geometrical motion in lung tumor located in lower lobe and adhered to chest wall and diaphragm on SBRT for NSCLC and PTV margin could be as large as 0.88, 0.98 and 2.68 cm in RL, AP and SI directions, respectively. However, a conventional 1.0 cm margin in the SI plane dose not suffice to compensate the geometrical variety of the tumor adhere to chest wall and diaphragm.

Abbreviations

4DCT:

Four-dimensional computed tomography

CBCT:

Cone-beam Computer Tomography

CTV:

Clinical target volume

EPID:

Electronic portal imaging device

GTV:

Gross target volume

ITV:

Internal clinical target volume

NSCLC:

Non-small cell lung cancer

PTV:

Planning target volume

SBRT:

Stereotactic Body Radiotherapy

TP-3DCT:

Treatment planning three-dimensional computed tomography

References

  1. Lanni Jr TB, Grills IS, Kestin LL, Robertson JM. Stereotactic radiotherapy reduces treatment cost while improving overall survival and local control over standard fractionated radiation therapy for medically inoperable non-small-cell lung cancer. Am J Clin Oncol. 2011;34(5):494–8. doi:10.1097/COC.0b013e3181ec63ae. Epub 2010/09/02.

    Article  PubMed  Google Scholar 

  2. Curran Jr WJ, Paulus R, Langer CJ, Komaki R, Lee JS, Hauser S, et al. Sequential vs. concurrent chemoradiation for stage III non-small cell lung cancer: randomized phase III trial RTOG 9410. J Natl Cancer Inst. 2011;103(19):1452–60. doi:10.1093/jnci/djr325. PubMed PMID: 21903745, PubMed Central PMCID: PMC3186782, Epub 2011/09/10.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  3. Borst GR, Sonke JJ, Betgen A, Remeijer P, van Herk M, Lebesque JV. Kilo-voltage cone-beam computed tomography setup measurements for lung cancer patients; first clinical results and comparison with electronic portal-imaging device. Int J Radiat Oncol Biol Phys. 2007;68(2):555–61. doi:10.1016/j.ijrobp.2007.01.014. Epub 2007/04/03.

    Article  PubMed  Google Scholar 

  4. de Boer HC, van Sornsen de Koste JR, Senan S, Visser AG, Heijmen BJ. Analysis and reduction of 3D systematic and random setup errors during the simulation and treatment of lung cancer patients with CT-based external beam radiotherapy dose planning. Int J Radiat Oncol Biol Phys. 2001;49(3):857–68. Epub 2001/02/15.

    Article  PubMed  Google Scholar 

  5. Abbas G, Pennathur A, Landreneau RJ, Luketich JD. Radiofrequency and microwave ablation of lung tumors. J Surg Oncol. 2009;100(8):645–50. doi:10.1002/jso.21334. Epub 2009/12/18.

    Article  PubMed  Google Scholar 

  6. Chang J, Mageras GS, Yorke E, De Arruda F, Sillanpaa J, Rosenzweig KE, et al. Observation of interfractional variations in lung tumor position using respiratory gated and ungated megavoltage cone-beam computed tomography. Int J Radiat Oncol Biol Phys. 2007;67(5):1548–58. doi:10.1016/j.ijrobp.2006.11.055. PubMed PMID: 17394950, PubMed Central PMCID: PMC2278042, Epub 2007/03/31.

    Article  PubMed  PubMed Central  Google Scholar 

  7. Hugo GD, Yan D, Liang J. Population and patient-specific target margins for 4D adaptive radiotherapy to account for intra- and inter-fraction variation in lung tumour position. Phys Med Biol. 2007;52(1):257–74. doi:10.1088/0031-9155/52/1/017. Epub 2006/12/22.

    Article  PubMed  Google Scholar 

  8. Sonke JJ, Lebesque J, van Herk M. Variability of four-dimensional computed tomography patient models. Int J Radiat Oncol Biol Phys. 2008;70(2):590–8. doi:10.1016/j.ijrobp.2007.08.067. Epub 2007/11/27.

    Article  PubMed  Google Scholar 

  9. Seppenwoolde Y, Shirato H, Kitamura K, Shimizu S, van Herk M, Lebesque JV, et al. Precise and real-time measurement of 3D tumor motion in lung due to breathing and heartbeat, measured during radiotherapy. Int J Radiat Oncol Biol Phys. 2002;53(4):822–34. Epub 2002/07/04.

    Article  PubMed  Google Scholar 

  10. Britton KR, Starkschall G, Tucker SL, Pan T, Nelson C, Chang JY, et al. Assessment of gross tumor volume regression and motion changes during radiotherapy for non-small-cell lung cancer as measured by four-dimensional computed tomography. Int J Radiat Oncol Biol Phys. 2007;68(4):1036–46. doi:10.1016/j.ijrobp.2007.01.021. Epub 2007/03/24.

    Article  PubMed  Google Scholar 

  11. Donnelly ED, Parikh PJ, Lu W, Zhao T, Lechleiter K, Nystrom M, et al. Assessment of intrafraction mediastinal and hilar lymph node movement and comparison to lung tumor motion using four-dimensional CT. Int J Radiat Oncol Biol Phys. 2007;69(2):580–8.

    Article  PubMed  PubMed Central  Google Scholar 

  12. Gauthier JF, Varfalvy N, Tremblay D, Cyr MF, Archambault L. Characterization of lung tumors motion baseline using cone-beam computed tomography. Med Phys. 2012;39(11):7062–70. doi:10.1118/1.4762563. Epub 2012/11/07.

    Article  PubMed  Google Scholar 

  13. Ecclestone G, Bissonnette JP, Heath E. Experimental validation of the van Herk margin formula for lung radiation therapy. Med Phys. 2013;40(11):111721. doi:10.1118/1.4824927. Epub 2013/12/11.

    Article  PubMed  Google Scholar 

  14. Hof H, Rhein B, Haering P, Kopp-Schneider A, Debus J, Herfarth K. 4D-CT-based target volume definition in stereotactic radiotherapy of lung tumours: comparison with a conventional technique using individual margins. Radiother Oncol. 2009;93(3):419–23. doi:10.1016/j.radonc.2009.08.040. Epub 2009/09/29.

    Article  PubMed  Google Scholar 

  15. Bradley JD, Nofal AN, El Naqa IM, Lu W, Liu J, Hubenschmidt J, et al. Comparison of helical, maximum intensity projection (MIP), and averaged intensity (AI) 4D CT imaging for stereotactic body radiation therapy (SBRT) planning in lung cancer. Radiother Oncol. 2006;81(3):264–8. doi:10.1016/j.radonc.2006.10.009. Epub 2006/11/23.

    Article  PubMed  Google Scholar 

  16. Ueda Y, Miyazaki M, Nishiyama K, Suzuki O, Tsujii K, Miyagi K. Craniocaudal safety margin calculation based on interfractional changes in tumor motion in lung SBRT assessed with an EPID in cine mode. Int J Radiat Oncol Biol Phys. 2012;83(3):1064–9. doi:10.1016/j.ijrobp.2011.07.043. Epub 2012/01/17.

    Article  PubMed  Google Scholar 

  17. Bissonnette JP, Franks KN, Purdie TG, Moseley DJ, Sonke JJ, Jaffray DA, et al. Quantifying interfraction and intrafraction tumor motion in lung stereotactic body radiotherapy using respiration-correlated cone beam computed tomography. Int J Radiat Oncol Biol Phys. 2009;75(3):688–95. doi:10.1016/j.ijrobp.2008.11.066. Epub 2009/04/28.

    Article  PubMed  Google Scholar 

  18. Hugo GD, Liang J, Campbell J, Yan D. On-line target position localization in the presence of respiration: a comparison of two methods. Int J Radiat Oncol Biol Phys. 2007;69(5):1634–41. doi:10.1016/j.ijrobp.2007.08.023. PubMed PMID: 18029112, PubMed Central PMCID: PMC2170894, Epub 2007/11/22.

    Article  PubMed  PubMed Central  Google Scholar 

  19. Grills IS, Hugo G, Kestin LL, Galerani AP, Chao KK, Wloch J, et al. Image-guided radiotherapy via daily online cone-beam CT substantially reduces margin requirements for stereotactic lung radiotherapy. Int J Radiat Oncol Biol Phys. 2008;70(4):1045–56.

    Article  PubMed  Google Scholar 

  20. Corradetti MN, Mitra N, Bonner Millar LP, Byun J, Wan F, Apisarnthanarax S, et al. A moving target: image guidance for stereotactic body radiation therapy for early-stage non-small cell lung cancer. Practical Radiation Oncology. 2013;3(4):307–15. doi:10.1016/j.prro.2012.10.005. Epub 2014/03/29.

    Article  PubMed  Google Scholar 

  21. Cai W, Dhou S, Cifter F, Myronakis M, Hurwitz MH, Williams CL, et al. 4D cone beam CT-based dose assessment for SBRT lung cancer treatment. Phys Med Biol. 2016;61(2):554–68. doi:10.1088/0031-9155/61/2/554. Epub 2015/12/20.

    Article  CAS  PubMed  Google Scholar 

  22. Liu HW, Khan R, Nugent Z, Krobutschek K, Dunscombe P, Lau H. Factors influencing intrafractional target shifts in lung stereotactic body radiation therapy. Practical Radiation Oncology. 2014;4(1):e45–51. doi:10.1016/j.prro.2013.02.013. Epub 2014/03/14.

    Article  PubMed  Google Scholar 

  23. Wang L, Feigenberg S, Fan J, Jin L, Turaka A, Chen L, et al. Target repositional accuracy and PTV margin verification using three-dimensional cone-beam computed tomography (CBCT) in stereotactic body radiotherapy (SBRT) of lung cancers. J Appl Clin Med Phys. 2012;13(2):3708. doi:10.1120/jacmp.v13i2.3708. Epub 2012/03/10.

    PubMed  Google Scholar 

  24. Li W, Purdie TG, Taremi M, Fung S, Brade A, Cho BC, et al. Effect of immobilization and performance status on intrafraction motion for stereotactic lung radiotherapy: analysis of 133 patients. Int J Radiat Oncol Biol Phys. 2011;81(5):1568–75. doi:10.1016/j.ijrobp.2010.09.035. Epub 2010/11/16.

    Article  CAS  PubMed  Google Scholar 

  25. Rietzel E, Chen GT, Choi NC, Willet CG. Four-dimensional image-based treatment planning: target volume segmentation and dose calculation in the presence of respiratory motion. Int J Radiat Oncol Biol Phys. 2005;61(5):1535–50. doi:10.1016/j.ijrobp.2004.11.037. Epub 2005/04/09.

    Article  PubMed  Google Scholar 

  26. Stroom JC, de Boer HC, Huizenga H, Visser AG. Inclusion of geometrical uncertainties in radiotherapy treatment planning by means of coverage probability. Int J Radiat Oncol Biol Phys. 1999;43(4):905–19. Epub 1999/03/31.

    Article  CAS  PubMed  Google Scholar 

  27. van Herk M, Remeijer P, Rasch C, Lebesque JV. The probability of correct target dosage: dose-population histograms for deriving treatment margins in radiotherapy. Int J Radiat Oncol Biol Phys. 2000;47(4):1121–35. Epub 2000/06/23.

    Article  PubMed  Google Scholar 

  28. Parker BC, Shiu AS, Maor MH, Lang FF, Liu HH, White RA, et al. PTV margin determination in conformal SRT of intracranial lesions. J Appl Clin Med Phys. 2002;3(3):176–89. doi:10.1120/1.1474308. Epub 2002/07/23.

    Article  PubMed  Google Scholar 

  29. Stevens CW, Munden RF, Forster KM, Kelly JF, Liao Z, Starkschall G, et al. Respiratory-driven lung tumor motion is independent of tumor size, tumor location, and pulmonary function. Int J Radiat Oncol Biol Phys. 2001;51(1):62–8.

    Article  CAS  PubMed  Google Scholar 

  30. Guckenberger M, Meyer J, Wilbert J, Richter A, Baier K, Mueller G, et al. Intra-fractional uncertainties in cone-beam CT based image-guided radiotherapy (IGRT) of pulmonary tumors. Radiother Oncol. 2007;83(1):57–64. doi:10.1016/j.radonc.2007.01.012. Epub 2007/02/20.

    Article  PubMed  Google Scholar 

  31. Worm ES, Hansen AT, Petersen JB, Muren LP, Praestegaard LH, Hoyer M. Inter- and intrafractional localisation errors in cone-beam CT guided stereotactic radiation therapy of tumours in the liver and lung. Acta Oncol. 2010;49(7):1177–83. doi:10.3109/0284186X.2010.498435. Epub 2010/07/02.

    Article  PubMed  Google Scholar 

  32. Rietzel E, Liu AK, Doppke KP, Wolfgang JA, Chen AB, Chen GT, et al. Design of 4D treatment planning target volumes. Int J Radiat Oncol Biol Phys. 2006;66(1):287–95. doi:10.1016/j.ijrobp.2006.05.024. Epub 2006/08/15.

    Article  PubMed  Google Scholar 

  33. Mori S, Endo M, Komatsu S, Yashiro T, Kandatsu S, Baba M. Four-dimensional measurement of lung tumor displacement using 256-multi-slice CT-scanner. Lung Cancer. 2007;56(1):59–67. doi:10.1016/j.lungcan.2006.11.011. Epub 2006/12/13.

    Article  PubMed  Google Scholar 

  34. Descovich M, McGuinness C, Kannarunimit D, Chen J, Pinnaduwage D, Pouliot J, et al. Comparison between target margins derived from 4DCT scans and real-time tumor motion tracking: insights from lung tumor patients treated with robotic radiosurgery. Med Phys. 2015;42(3):1280–7. doi:10.1118/1.4907956. Epub 2015/03/05.

    Article  PubMed  Google Scholar 

  35. Dobashi S, Sugane T, Mori S, Asakura H, Yamamoto N, Kumagai M, et al. Intrafractional respiratory motion for charged particle lung therapy with immobilization assessed by four-dimensional computed tomography. J Radiat Res. 2011;52(1):96–102. Epub 2011/02/05.

    Article  PubMed  Google Scholar 

  36. Timmerman R, Papiez L, McGarry R, Likes L, DesRosiers C, Frost S, et al. Extracranial stereotactic radioablation: results of a phase I study in medically inoperable stage I non-small cell lung cancer. Chest. 2003;124(5):1946–55. Epub 2003/11/08.

    Article  PubMed  Google Scholar 

  37. Lax I, Blomgren H, Naslund I, Svanstrom R. Stereotactic radiotherapy of malignancies in the abdomen. Methodological aspects. Acta Oncol. 1994;33(6):677–83. Epub 1994/01/01.

    Article  CAS  PubMed  Google Scholar 

Download references

Acknowledgments

All authors in this study and many colleagues helped to collect and analyze data. Their support and helps should be appreciated.

Funding

This study was supported in part by institutional science foundation of the first affiliated hospital of Xi’an Jioatong University (2013YK27).

Availability of data and materials

All data and materials have been presented in the manuscript.

Authors’ contributions

X-ZZ conceived of the study, and participated in its design and coordination. YL design and carried out the studies, participated in the sequence alignment and drafted the manuscript. J-LM help to draft the manuscript and contributed to the writing of the manuscript, XC and F-WT help to carried out the studies. All authors read and approved the final manuscript.

Competing interests

The authors declare that they have no competing interests.

Consent for publication

Not applicable.

Ethics approval and consent to participate

The study was approved by the Institutional Committee for Clinical Research and the Local Ethics Committee in The First Affiliated Hospital, Xi’an Jiaotong University of medical college. Eligible patients were given written informed consent.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Xiao-zhi Zhang.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Li, Y., Ma, Jl., Chen, X. et al. 4DCT and CBCT based PTV margin in Stereotactic Body Radiotherapy(SBRT) of non-small cell lung tumor adhered to chest wall or diaphragm. Radiat Oncol 11, 152 (2016). https://doi.org/10.1186/s13014-016-0724-5

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13014-016-0724-5

Keywords